Service \ Assembly & integration \ Reagent storage & lyophilization

REAGENT STORAGE & LYOPHILIZATION

On-board reagent storage — blisters, capsules, lyophilized beads and in-situ drying — loaded on our own lines.

MATERIAL

Blister · capsule · lyophilized bead · in-situ

MINIMUM ORDER

No strict minimum

LEAD TIME

3–4 weeks from agreed requirements

\ OVERVIEW

Reagent storage & lyophilization

Getting reagent into a microfluidic device and keeping it stable there is a separate engineering problem from making the device. Four formats cover it. For liquid: blisters and capsules — each a reservoir or chamber inside the cartridge, sized to the reagent volume and the release mechanism. For dry: lyophilized beads, metered and placed as discrete units, and in-situ lyophilization, where the reagent is dispensed into the device cavity and dried in place. All four are loaded on our own assembly lines, before closure; standalone beads also ship in customized storage tubes.

hero.jpeg
v19_01_four_formats_transparent.png

4 formats

blister · capsule · bead · in-situ

v19_02_no_cold_chain_transparent.png

No cold chain

the point of drying the reagent

v19_03_no_strict_moq_transparent.png

No strict MOQ

pricing improves with batch size

v19_04_three_four_weeks_transparent.png

3–4 weeks

from agreed requirements

v2_05_iso-13485-certified.png

ISO 13485

certified · FDA registered​

\ CAPABILITIES

Capabilities and technical specifications

Reagents are loaded on the line that builds the device — treatment before closure, reagent verified before it is sealed in — in whichever dried form the cavity calls for, a metered bead or in-situ drying. Your formulation stays yours; we develop the drying cycle around it.

Capability

Dose, residual-moisture and shelf-life figures are agreed per formulation at DFM.

PARAMETER

SPECIFICATION

Formats

Blister · capsule · liquid storage tube · lyophilized bead · in-situ lyophilization

Customization

Cavity, blister, capsule and tube geometry designed to the reagent volume and to how the chip uses it

Dried forms

Discrete spherical beads, or an in-cavity cake dried in the device

Cycle development

Freeze, primary-dry and secondary-dry cycle built around your formulation to stay below its collapse temperature

Loading

Filling, bead detection and placement on our assembly lines, before device closure

Packaging

Dedicated reagent tubes with optional desiccant; loaded devices ship assembled and packaged

Bead volume and dose CV · residual moisture · reconstitution time · shelf life

Not yet published

Quality system

ISO 13485 certified · FDA registered

Where we stop — the division of work

AREA

HICOMP

YOU

Reagent formulation

—

Your assay chemistry, concentrations, buffer and lyoprotectant system

Storage hardware

Blister or capsule, designed to your volume and release mechanism

Volume, actuation method, contact-material constraints

Lyophilization

Cycle development, then drying as a discrete bead or in situ in the device cavity

Formulation, and the acceptance criteria for activity after drying

Loading

Filling, bead detection and placement, in the correct sequence and before closure

—

Device

Chip, cartridge, bonding, treatment, assembly, packaging

—

We do not want your formulation and we do not need it.

You keep the chemistry and the IP; we build the device and the process that holds it. Where formulation and cycle are incompatible, we say which part is the problem rather than quietly changing one of them.

Equipment

Two kinds of machine do most of the work. The lyophilizers run the drying cycle; the micro dispensers meter reagent — into a droplet for beads, or straight into the cavity for in-situ drying.

hero.jpeg

Production lyophilizer — the freeze, primary-dry and secondary-dry cycle runs here

hero.jpeg

Micro dispenser — the same metering step feeds both dried forms

hero.jpeg

Device trays on the shelf — reagent dried in the cavity, on the part

hero.jpeg

Bead detection and placement — each bead detected, then placed before closure​​​

\ MATERIALS

Materials

Two liquid formats, two dried forms and the storage tube. The choice is set by volume, release point in the fluidic sequence, and whether the reagent has to be ambient-stable.

Storage formats

FORMAT

HOW THE REAGENT IS HELD

RELEASED BY

SUITS

Blister

Liquid sealed between a formed cavity and a foil laminate lid

Mechanical actuation bursts the seal at a designed weak point

Single-use cartridges with a fixed dose and a defined actuation stroke

Capsule

Liquid in a reservoir or chamber inside the cartridge, sealed until use

Pierced, crushed or opened on actuation

Larger volumes, reagents that must stay isolated from the device until use, and release sequences a blister cannot give

Liquid storage tube

Liquid in a dedicated tube customized for the reagent volume and the way the chip uses it

The chip's method of use, which the tube is customized for

Reagents whose quality and concentration must stay stable in the chip

Lyophilized bead

Dried to a discrete solid sphere, then placed into the device

Reconstituted by the sample or by an on-board buffer

Ambient storage, multi-reagent devices, and anywhere the dose has to be repeatable unit to unit

In-situ lyophilization

Dispensed into the device cavity as liquid and dried in place

Reconstituted by the sample or by an on-board buffer

Cavities a bead will not fit or reach, larger dried volumes, and geometries where a placed bead would move in transit

Materials that touch the reagent

MATERIAL

WHERE IT APPEARS

WHAT IT HAS TO DO

Foil laminate

Blister lidding, pouch packaging

Moisture and oxygen barrier — the shelf-life claim is usually limited by this laminate, not by the chemistry

PP, PE, COC/COP

Blister cavities, capsule bodies, dried-reagent cavities, chip bodies

Low extractables, low sorption of the analyte or the label, and compatibility with the sealing method

Sealing film and PSA

Chamber closure, secondary seals

Seal integrity through shipping, temperature cycling and the actuation event

Desiccant

Packaging of lyophilized product

Holds residual moisture down through storage — supplied as an option with reagent-tube packaging

The buffer that works at room temperature can destroy the product during freezing.

Sodium phosphate is a standard bench choice, but on freezing the disodium salt crystallizes preferentially and the residual solution swings acidic by several pH units, exactly where the protein is. Formulations that survive this have usually moved to histidine or potassium phosphate; Tris is not the fix, since its pKa shifts roughly 0.03 units per degree. It is your formulation to set, but worth checking before we develop a cycle around it.

Drying below the collapse temperature is what buys shelf life.

Product temperature above collapse during primary drying gives a shrunken, glassy cake instead of a porous one. It can still pass a visual release check, but it reconstitutes more slowly, retains more residual moisture and loses activity over months — so it fails stability testing. Cycle development is largely the work of finding that temperature for your formulation and staying under it.

Lyophilized bead vs. in-situ lyophilization

Neither is the default; the cavity usually decides. Grade-level data on the contact materials is in the Materials guide.

ASPECT

LYOPHILIZED BEAD

IN-SITU LYOPHILIZATION

Where drying happens

As a free-standing droplet, before the device is involved

Inside the device cavity, on the part

Dose control

Metered as a discrete unit and checked before placement, so a bad unit is rejected rather than sealed in

Set by dispensing accuracy into the cavity; verified on the part

Cavity constraints

The cavity has to accept and retain a formed bead

No shape constraint — the cake takes the cavity's shape

Volume

Best suited to small, repeatable doses

Handles larger dried volumes

Movement in transit

A placed bead can move unless retained

Cake is formed in place and stays there

Rework

A rejected bead costs a bead

A failed dry costs the part it is in

\ PROCESS

Lyophilization process

Two ways to dry, chosen by the cavity; one closure and packaging line for both.

  • Track A — Lyophilized beads

Dose fixed before drying; every bead checked before the device is involved.

hero.jpeg

Formulation and cycle development

Your formulation received with the acceptance criteria for activity after drying; a freeze, primary-dry and secondary-dry cycle built to stay below its collapse temperature.

Below collapse temperature

hero.jpeg

Metering

Droplets metered as discrete units, so dose is fixed before drying.​

​

​

​

​

​

​

Dose fixed before drying

hero.jpeg

Freezing

Frozen as discrete spheres. Freezing sets the ice-crystal structure, and so the pore structure of the finished bead.

​

​

​

Freezing rate sets pore structure

hero.jpeg

Primary and secondary drying

Sublimation under vacuum with product temperature held below collapse, then desorption of bound water to the residual-moisture target.

​

Residual-moisture target

hero.jpeg

Inspect and load

Each bead detected and checked, then placed into the device before closure. A missing or damaged bead is rejected, not sealed in.

​

​​

​

Detected before placed

  • Track B — In-situ lyophilization

Dried in the device cavity, so the cake takes the cavity’s shape and cannot move in transit.

hero.jpeg

Cavity review and cycle development

Cavity volume, geometry and material checked against the dried volume and the reconstitution path; the cycle developed as in Track A.​

Cavity sets the cake

hero.jpeg

Dispensing into the cavity

Liquid reagent metered directly into the device cavity, on the line. Dose accuracy is now dispensing accuracy.

​

​

Verified on the part

hero.jpeg

Freezing in place

Frozen in the cavity. The cavity sets the geometry, so freezing rate across the part is the parameter that has to be uniform.

Uniform freezing across the part

hero.jpeg

Drying and cake inspection

Same cycle discipline as bead work. Cake inspected in the cavity: appearance, fill position, no collapse.

​

A failed dry costs the part

  • Closure and packaging — common to both tracks

hero.jpeg

Closure and sealing

The device closed — PSA lamination or film seal — without the dried reagent seeing ambient humidity; blister and chamber seals inspected on the line.

​

No ambient exposure before closure

hero.jpeg

Barrier packaging

Standalone beads in dedicated reagent tubes with optional desiccant; loaded devices in moisture-barrier packaging where required.

​

Foil laminate · optional desiccant

hero.jpeg

Stability handover

Released with the controlled inputs to your stability study — residual moisture, cake structure, seal integrity and barrier packaging. Shelf life is established by your study, not asserted by us.

Your stability study

  • Blisters and capsules — fill and seal

The liquid formats take a shorter route, on the same line and in the same sequence relative to treatment and closure.

Requirement

Reagent volume, release mechanism, actuation force and contact-material constraints.

​

​

​

Format and geometry

Blister cavity or capsule chamber designed against the cartridge interface, not selected from a catalog.​​​​​​

Tooling and samples

Cavity tooled and sampled for fit, fill, seal and release.

​

​

​

​

Fill and seal

Dosed and sealed on the line, before device closure.

​

​

​

​

Integration

Assembled into the device and inspected.

​

​

​

​

​

hero.jpeg

\ QUALITY CONTROL

Quality control

Bead verification

Presence and condition detected automatically before placement, so a missing or damaged bead is rejected rather than sealed into a finished device

Seal integrity

Blister and chamber seals inspected on the line

​

Analytical release

Residual moisture (Karl Fischer or loss on drying), reconstitution time, activity against the customer’s assay

Quality system

ISO 13485 certified · FDA registered

In-cavity cake

Inspected in the device before closure — cake appearance, fill position and absence of collapse

​

Visual inspection

100% automated vision inspection through the assembly line, with defect mapping

Packaging

Reagent tubes with optional desiccant; moisture-barrier packaging where required

​

Residual moisture is usually a better predictor of shelf life than day-one activity.

A dried product carrying too much residual water loses potency slowly over months — it passes release and fails at twelve months. The target is a window, not a minimum, since over-drying destabilizes some proteins; agree a residual-moisture specification at the start, not a potency specification alone.

\ LEAD TIME & MOQ

Lead time & MOQ

v20_01_lead_time_req_transparent.png

Lead time

3 to 4 weeks from receipt of specific requirements — the formulation, the volume, the release mechanism and the acceptance criteria

v20_04_delivery_beads_tube_transparent.png

Delivery format

Standalone beads packaged in dedicated reagent tubes, with optional desiccant. In-situ dried reagent ships inside the device

v20_02_cycle_development_transparent.png

Cycle development

For a formulation that has never been dried before, it sits ahead of that clock and is quoted separately

​

v20_05_integrated_delivery_transparent.png

Integrated delivery

Where the reagent is loaded into a device we build, it ships as part of the assembled, packaged unit

​

v20_03_min_order_batch_transparent.png

Minimum order

No strict minimum — pricing improves with batch size. Development quantities and production batches are both normal

v20_06_pricing_per_program_transparent.png

Indicative pricing

Quoted per program

​

​

 

\ SHOWCASE

Showcase

Sample parts

The formats, photographed on parts made here.

hero.jpeg

Reagent storage blisters

liquid held against a burstable foil seal, released by a designed actuation stroke

hero.jpeg

Reagent capsule

a sealed reservoir inside the cartridge, isolated from the device until the moment of use

hero.jpeg

Lyophilized beads

metered, dried and checked before the device is involved, then placed robotically, or shipped standalone in dedicated reagent tubes

hero.jpeg

In-situ lyophilization

the cake takes the shape of the cavity and cannot move in transit

Applications

hero.jpeg

AMBIENT-STABLE CARTRIDGES

A cartridge that needs no refrigeration is a different commercial product — drying the reagent removes the cold chain.

hero.jpeg

SAMPLE PREPARATION

Multi-reagent cartridges mixing blisters, capsules and dried reagent in one fluidic sequence.

​

hero.jpeg

SAMPLE-TO-ANSWER

PCR and lysis reagents stored in the cartridge, reconstituted by the sample or an on-board buffer.

​

hero.jpeg

LIBRARY PREPARATION

Enzymatic reagents dried below their collapse temperature so activity survives storage, not just release.

FAQ

\ RELATED

Related services

hero.jpeg

The lines that load, seal and inspect the reagent — the same lines that build the device.

Bonding fixture press station

Sealing, PSA lamination and chamber closure without heat the reagent cannot take.

hero.jpeg

Wettability control applied in line, before reagent loading and the PSA closure.

Abstract White Waves

\ START HERE

TELL US ABOUT YOUR PROJECT

Reply within 2 business days.
NDA on request.

Project stage
Prototype
Pilot batch / process verification
Production
Quantity (per month at production)
Fewer than 20
20–200
200–2,000
More than 2,000
How did you hear about us?

By sending this, you agree that we may contact you about your inquiry. We do not add you to a mailing list.

Prefer to talk it through first? Book a design review →

bottom of page